While a patient is being ventilated with NeoFlow on a Dräger E-4, an audible and visual alarm is active, and all breaths are being time-triggered. What is the most appropriate action for the res-piratory therapist to take?
a. Suction the endo-tracheal tube c. Dry out the flow sensor
b. Change the venti-lator out d. Switch to Au-toFlow
ANS: C
If the NeoFlow sensor becomes partially occluded with water (or secretions), it will cease opera-tion and trigger audible and visual alarms. When this occurs, the ventilator will continue to oper-ate in the pressure-control mode.