Explain the potential problem with heated wire circuits inside an inspiratory tubing.

What will be an ideal response?

One potential problem with this system is found when the distal temperature probe is placed at the patient connection inside a heated incubator that is set at a higher temperature than the humidifier. As the inspiratory tubing enters the incubator, the gas is heated to the set temperature of the incubator environment. This causes the temperature probe to sense the higher temperature and shut down the heater wires. The result is a buildup of condensation in the inspiratory tubing.

Health Professions

You might also like to view...

The exact time at which a patient started to experience signs and symptoms of a possible stroke

is important. Indicate whether the statement is true or false

Health Professions

Preeclampsia MOST commonly occurs after the ____ week of gestation.

A. 12th B. 20th C. 24th D. 30th

Health Professions