The operational verification procedure (OVP) involves checking the ventilator circuit for leaks. Ventilator settings that could be used to perform this procedure include which of the following?

a. Tidal volume (VT) = 500 mL, Flow rate = 60 L/min, High pressure limit = 50 cm H2O
b. VT = 1000 mL, Flow rate = 20 L/min, High pressure limit = maximum
c. VT = 500 mL, Flow rate = 20 L/min, High pressure limit = maximum, Inspiratory pause = 2 seconds
d. VT = 200 mL, Flow rate = Maximum, High pressure limit = 50 cm H2O, Inspira-tory pause = 1 second

ANS: C
To check for leaks in the ventilator circuit the operator should set the tidal volume to 500 mL, the gas flow low (e.g., 20 L/min), the maximum pressure limit high (e.g., 100-120 cm H2O), and an inspiratory pause of 1-2 seconds.

Health Professions

You might also like to view...

Which of the following statements is correct?

A. Dog bite wounds are more likely to become infected than human bite wounds. B. Human bite wounds are more likely to become infected than dog bite wounds. C. Only human bite wounds that occur on the dorsal hand are likely to become infected because the skin on the dorsum of the hand is extremely thin. D. All bite wounds will become infected and therefore prophylactic antibiotics should be prescribed by a physician.