During calibration of a Clark polarographic O2 analyzer, you cannot get the sensor to read 100% when exposed to 100% O2, even after adjusting its calibration control. The unit has new batteries

Which of the following actions would be correct?
a. Use it only with low FiO2 values.
b. Send the device out for repair.
c. Replace the unit's batteries.
d. Change the sensor or electrode.

ANS: D
If the analyzer still does not calibrate on fresh batteries, the problem is probably a depleted sen-sor. With most analyzers, a depleted sensor must be replaced (some Clark electrodes can be re-charged).

Health Professions

You might also like to view...

A 35-year-old male complains of a "pressure"-type headache to the back of his head and neck. He states that he woke up with the discomfort this morning, and the pain has steadily gotten worse through the day. He denies any radiation of the pain and denies photophobia but states that he is slightly nauseated. Based on this clinical presentation, the paramedic should be suspicious of what type of