During the transport of a patient being ventilated with a Newport HT50 ventilator, the low air-way pressure/apnea indicator begins to blink red and the "check proximal line" alarm is triggered. What is the most appropriate action?

a. Suction the pa-tient's airway. c. Check the circuit's integrity.
b. Increase the pres-sure limit. d. Switch the mode of ventilation.

ANS: C
The "check proximal line" alarm occurs during inspiration when the proximal pressure measured is significantly different from the internal backup pressure sensor measurement taken inside the ventilator. In addition to the message, the low airway pressure/apnea indicator blinks red. Causes include disconnections, kinking of the circuit, and a water-filled proximal sensor line.

Health Professions

You might also like to view...

What is the typical shape of a scabies burrow lesion?

A) S-shaped B) Circular C) Square D) Half-moon

Health Professions

One of the biggest challenges to an administrator or manager is keeping employees motivated. Which of the following is a proven way to "motivate" employees?

A) ?Recognize an employee for accomplishing goals within an organization, team, and where management encourages innovative and experimental ideas. B) Foster pride in "ownership" of the job or assignment; make the employee feels invested in the position. C) Use Specific, Measurable, Attainable, Results-oriented, and Time-bound guidelines to reward an employee with cash or time credit bonuses. D) All of the above