The respiratory therapist observes the following pressure-time and flow-time scalars following a patient being intubated and placed on a mechanical ventilator using volume ventilation. The most appropriate action is which of the following?
a. Increase the set flow rate.
b. Switch to pressure control.
c. Increase the set tidal volume.
d. Change to a decelerating flow pattern.
ANS: A
When the flow rate setting is inadequate, the pressure-time scalar takes on a concave shape dur-ing inspiration when the patient is actively inhaling. In this case the demand exceeds the ventila-tor setting and capability. The inspiratory demand of the patient will be provided for by increas-ing the set flow.
You might also like to view...
The Pharyngeo-tracheal Lumen (PtL) airway is most similar in design and function as what other airway device?
A) Esophageal Tracheal Combitube B) King LT airway C) Oropharyngeal airway D) Endotracheal tube airway
The placement of the patella in relationship to the femorotibial joint space on an AP knee projec-tion is affected by