When the Dräger Babylog 8000+ is used in the pressure support mode, the patient's peak inspir-atory flow is 30 L/min. At what flow rate will the ventilator cycle to expiration?

a. 3.0 L/min c. 6.0 L/min
b. 4.5 L/min d. 7.0 L/min

ANS: B
Pressure support breaths are flow-cycled on the Dräger Babylog 8000+ at 15% of peak flow.